Abstract for rohling_bmvc

Proceedings of the 1996 British Machine Vision Conference, Edinburgh

CORRECTING MOTION-INDUCED REGISTRATION ERRORS IN 3-D ULTRASOUND IMAGES

Robert Rohling and Andrew Gee

September 1996

3-D ultrasound is a new medical imaging technique which can quickly and non-invasively deliver clinically useful renderings of internal structures. The technique is vulnerable to tissue motion during the scan which can result in an inconsistent 3-D data set. This paper describes an algorithm which corrects for tissue motion and other errors that cause mis-registration during a scan. The resulting 3-D data sets are clearer and of enhanced diagnostic utility. The registration is achieved by applying visual tracking techniques outside their traditional area of application.The algorithm is automatic, incremental and easily incorporated into existing 3-D ultrasound systems.


(ftp:) rohling_bmvc.ps.gz (http:) rohling_bmvc.ps.gz
PDF (automatically generated from original PostScript document - may be badly aliased on screen):
  (ftp:) rohling_bmvc.pdf | (http:) rohling_bmvc.pdf

If you have difficulty viewing files that end '.gz', which are gzip compressed, then you may be able to find tools to uncompress them at the gzip web site.

If you have difficulty viewing files that are in PostScript, (ending '.ps' or '.ps.gz'), then you may be able to find tools to view them at the gsview web site.

We have attempted to provide automatically generated PDF copies of documents for which only PostScript versions have previously been available. These are clearly marked in the database - due to the nature of the automatic conversion process, they are likely to be badly aliased when viewed at default resolution on screen by acroread.